... | ... | @@ -143,11 +143,7 @@ Use the yellow marker pen slide with a coverslip (on the desk next to the lens c |
|
|
8. Go to ‘Maintain’, and drag out the ‘Dual Camera Adapter Alignment’ tool
|
|
|
9. Change the ‘Rot [°]’ value. Store current position in the ‘Dual Camera Alignment’ tool
|
|
|
10. Take another Snap of the marker sample and measure the angle again. Repeat until you measure a value lower than 0.05°. Store current position in the ‘Dual Camera Alignment’ tool
|
|
|
11. Tick 'Calibration pattern' in the ‘Dual Camera Adapter Alignment’ tool and check the rotation again
|
|
|
|
|
|

|
|
|
|
|
|
12. Untick 'Calibration pattern'
|
|
|
|
|
|
|
|
|
</details>
|
... | ... | |