Update Zeiss Elyra 7 authored by Clara Bodner's avatar Clara Bodner
...@@ -176,14 +176,6 @@ Use the yellow marker pen slide with a coverslip (on the desk next to the lens c ...@@ -176,14 +176,6 @@ Use the yellow marker pen slide with a coverslip (on the desk next to the lens c
![cam_x_lineprofile](uploads/0b92f0c052b16846ceaaf2badf438694/cam_x_lineprofile.PNG) ![cam_x_lineprofile](uploads/0b92f0c052b16846ceaaf2badf438694/cam_x_lineprofile.PNG)
14. Tick 'Calibration pattern' and check XY overlap, fine-tune if necessary. Store current position!
![calibration_pattern](uploads/82161269165ec234399dc008455ae293/calibration_pattern.PNG)
15. Also check rotation again with the 'Calibration pattern'. Especially the dots in the center should be well aligned
![alignment_perfect](uploads/aa812ef246b98650bcd219c65a3c7b2f/alignment_perfect.PNG)
</details> </details>
<details> <details>
... ...
......