Update Zeiss Elyra 7 authored by Lukas Velas's avatar Lukas Velas
......@@ -206,10 +206,16 @@ Procedure:
![image](uploads/728fc21c8407920d336f7f1e6c5524b3/image.png){width=404 height=275}
- Select the following parameters and click Apply, save the Channel Alignment file.
- Select the following parameters and click Apply, save the Channel Alignment *.xml file.
![image](uploads/013bc4dd4c68164e185734f6ecc2ab18/image.png){width=411 height=455}
- Now, select a SIM processed image of your sample as in Input in the Channel Alignment (Extended) method
- Tick "Load Transformation" and load the previously saved *.xml channel alignment file. Click Apply.
- A new channel-corrected image is produced with a suffix - Channel Alignment (Extended). Save this file.
![image](uploads/78736917ea4fc54667ba132ed8b0fbf3/image.png){width=411 height=358}
</details>
......
......