@@ -130,12 +130,14 @@ Example of a Track2 with 405 nm laser activated, using DL QUAD 1 SBS LP 560 beam
...
@@ -130,12 +130,14 @@ Example of a Track2 with 405 nm laser activated, using DL QUAD 1 SBS LP 560 beam
-
In the Acquisition Tab, Imaging Setup window, add SMLM track bz clicking +SMLM from the drop down window.
-
In the Acquisition Tab, Imaging Setup window, add SMLM track bz clicking +SMLM from the drop down window.
-
Set up your imaging path:
-
Set up your imaging path:
-
1) Choose laser,
1) Choose laser,
-
2) TIRF for the field of view lens,
2) choose TIRF for the field of view lens,
-
3) Illumination mode (EPI/HILO/TIRF),
3) set illumination mode to EPI/HILO/TIRF,
-
4) don't change the Collimator,
4) don't change the Collimator,
-
5) FS LBF 405/488/561/642 beam splitter,
5) choose FS LBF 405/488/561/642 beam splitter,
-
6) use 3D SMLM slider onlz for 3D SMLM with double helix PSF,
6) use 3D SMLM slider only for 3D SMLM with double helix PSF,
7) pick 100x 1.46NA or 63x 1.46NA objective
8) on the emission side, choose 1x Tube lens and a suitable DuoLink beam splitter
[
Zeiss_Elyra_Setting_up_TIRF_imaging.pdf
](