Update Zeiss Elyra 7 authored by Lukas Velas's avatar Lukas Velas
......@@ -79,7 +79,7 @@ Always make sure that the system is back at 28.5°C and the CO<sub>2</sub> is of
</summary>
- Move to Acquisition Tab
- Move to the Acquisition Tab
- In the Imaging Setup window click on +Lattice SIM: This will create a Track1, if you need more tracks click additionally on +Lattice SIM
- Activate respective lasers (405, 488, 561 and 642 nm are available)
- In the SIM Grating changer 'Optimal Grating' should be checked
......@@ -124,7 +124,7 @@ Example of a Track2 with 405 nm laser activated, using DL QUAD 1 SBS LP 560 beam
<summary>
### `Setting up TIRF`
### `Setting up TIRF (old)`
</summary>
......@@ -144,7 +144,7 @@ Example of a Track2 with 405 nm laser activated, using DL QUAD 1 SBS LP 560 beam
</details>
## Dual-Camera Alignment & Colocalization Analysis
## Dual-Camera Alignment & Colocalization Analysis (old)
Camera Alignment has to be checked/done before **each experiment** using both cameras and for **each used filter** in the camera link and for **each used objective** individually. It has to be done again if the camera was pushed by accident, or if the cover of the camera link has been opened.
......@@ -152,7 +152,7 @@ Camera Alignment has to be checked/done before **each experiment** using both ca
<summary>
### `Camera Rotation`
### `Camera Rotation (old)`
</summary>
......@@ -189,7 +189,7 @@ Use the yellow marker pen slide with a coverslip (on the desk next to the lens c
<summary>
### `Camera xyz-Alignment`
### `Camera xyz-Alignment (old)`
</summary>
......@@ -219,7 +219,7 @@ Use the yellow marker pen slide with a coverslip (on the desk next to the lens c
<summary>
### `Channel Alignment`
### `Channel Alignment (old)`
</summary>
......@@ -274,7 +274,7 @@ Channel Alignment has to be done before **each colocalization experiment** using
<summary>
### `Data Handling and Transfer`
### `Data Handling and Transfer (old)`
</summary>
......
......