@@ -191,6 +191,17 @@ Channel Alignment consists of three steps:
2.
Finding the coordinate transformation between the two channels. This is a post processing step and can be done after your measurement.
3.
Applying the coordinate transformation to your recorded images. This is also a post processing step applied during your image analysis.
Procedure:
-
Mount TetraSpeck multicolor beads. Use the same objective as for your sample. Look for a spot with many evenly distributed beads.

{width=1065 height=1065}
-
Record a 5μm large Z stack around the beads using the same Tracks as for imaging your sample.

{width=408 height=405}