Update Zeiss Elyra 7 authored by Lukas Velas's avatar Lukas Velas
<img align="right" width="100" height="73" src="uploads/9cec1f9af6048a9fafc32567a125911b/MPL-Logo-RGB-Blau-2019-05-24.jpg">
[[_TOC_]]
# Profile
| ![Elyra](uploads/f53ec5c341f61133fa78aedb7630c39e/Elyra.png){width=5000} | Inverse super-resolution microscope for lattice SIM, 3D SMLM (PALM/STORM), TIRF and laser widefield imaging. Apotome mode for optical sectioning of thick specimen. Dual-camera port with 2 sCMOS cameras for fast acquisitions. Lasers to image dyes from blue to far red. Multi-positioning (Piezo-driven XY stage with Z-Piezo insert) and hardware autofocus. Environmental box for live imaging with temperature and CO2 control. Widefield illumination: Excelitas Xylis (white LED), halogen lamp. Processing PC for offline image processing (room 1.723) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
# Manuals & Guides
## Startup & Shutdown
<details>
<summary>
### `Startup`
</summary>
1. The Elyra microscope has only one switch that needs to be turned on (Main Switch), which is a turn knob. Do not turn any other switches on/off. Next, you will need to switch on the computer under the table.
![PXL_20240320_095753421](uploads/6243a65ae74f7d9c37bec3fa7fe80a35/PXL_20240320_095753421.jpg)
2. If at any point during your experiment a loud beeping sound goes off, first check the CO<sub>2</sub> sensor (and leave the room if the sound is coming from the sensor). If not, the sound is coming from the cooling system behind the Elyra and you need to [refill the cooling liquid](Maintenance/Elyra 7#Cooling-System).
3. Choose the correct objective you want to use for imaging (via the touch panel, menu point 'Microscope') and press 'Load position' to bring the objective down to its lowest position.
![PXL_20240320_100556693](uploads/944433d431b4d0027717cf627607037c/PXL_20240320_100556693.jpg){width=49%} ![PXL_20240320_100603288](uploads/7243fdb1c0c2344cdd05237a1b72ec56/PXL_20240320_100603288.jpg){width=49%}
4. Start the 'Zen black' Software and press 'Start up System'
5. The temperature inside the incubator will be by default at 27°C to minimize drift. Use the oil for 30°C. Immersion oil bottles for 30°C and 37°C are stored in the incubator at the bottom right.
![PXL_20240320_091559891](uploads/17793d9cf5156c1cff3710c2f8ec3da2/PXL_20240320_091559891.jpg){width=50%}
6. Put on your sample and focus on it using the 'Locate' tab in the Zen Software (press 'Fluorescence' or 'Transmitted').
7. With the front window open and the light in the incubator on (switch to the left, on the microscope table) you can safely navigate larger distances when using the joystick (which has an inverse logic).
For the front flap you will need to open (to move flap up) and tighten (to keep the flap in place) the little red screw. Please handle this window with care!
8. All doors on the incubator need to be properly closed before you can start imaging in the Acquisition tab as otherwise the interlock is not released and the lasers will not be active. You can check for this: if one of the LEDs on the main controller is blinking, one of the laser interlocks has not been recognized. Unfortunately, there is no feedback which interlock is affected, so you need to check them all and potentially, open and close all doors again or even lift the condenser and put it back down.
![PXL_20240320_091545142](uploads/771b95ca313e0a923254dbf26539bca8/PXL_20240320_091545142.jpg){width=50%} ![PXL_20240320_092901039](uploads/abfc3c39fba668f2f6dfff165d7c3a91/PXL_20240320_092901039.jpg){width=48%}
</details>
<details>
<summary>
### `Shutdown`
</summary>
- Close Zen Black
- Press the Load position on the touch panel
- Clean immersion objectives with lens cleaning paper
![Cleaning](uploads/93470e14dd699b18acd95bef9b1d8c97/Cleaning.jpg){width=40%}
Leaving system for the next user
- Logout from the Windows session
Full shutdown (last user of the day)
- Shut down the PC
- Switch off the microscope
Always make sure that the system is back at 27°C and the CO<sub>2</sub> is off when you have done any live cell imaging.
</details>
## Acquisition Setup
<details>
<summary>
### `Setting up Lattice SIM`
</summary>
[Zeiss_Elyra_Setting_up_Lattice_SIM.pdf](uploads/8a812e218450d11bab8caa13642c6510/Zeiss_Elyra_Setting_up_Lattice_SIM.pdf)
</details>
<details>
<summary>
### `Setting up TIRF`
</summary>
[Zeiss_Elyra_Setting_up_TIRF_imaging.pdf](uploads/d5abb8b971c53cbf7a21b318be59d1c3/Zeiss_Elyra_Setting_up_TIRF_imaging.pdf)
</details>
<details>
<summary>
### `Environmental Control`
</summary>
[Zeiss_Elyra_Environmental_controls_01.pdf](uploads/9e4a1088f421fd1742a2dd0076f3a89a/Zeiss_Elyra_Environmental_controls_01.pdf)
</details>
## Dual-Camera Alignment & Colocalization Analysis
Camera Alignment has to be checked/done before **each experiment** using both cameras and for **each used filter** in the camera link and for **each used objective** individually. It has to be done again if the camera was pushed by accident, or if the cover of the camera link has been opened.
<details>
<summary>
### `Camera Rotation`
</summary>
Use the yellow marker pen slide with a coverslip (on the desk next to the lens cleaning paper) and the slide with beads (multispec low, they are stored in the fridge close to the Elyra).
![PXL_20240321_131615341](uploads/9c0a3a7f70adb315ded4ebd63222571c/PXL_20240321_131615341.jpg){width=40%} ![PXL_20240321_131830944](uploads/8fb5c5277bf0106adaef969aefefa79d/PXL_20240321_131830944.jpg){width=40%}
1. Select the objective that you will use for imaging
2. Set up camera 2 (TV2) for calibration (TV1 is not calibrated): lattice SIM, LBF 405/488/561/642, LP560, use the 42µm grid
![cam_rotation_perfect_63x](uploads/80b10ebcf333aae8a5d4541459184fe9/cam_rotation_perfect_63x.PNG)
3. Focus on the lattice pattern with 'Continuous' and take a ‘Snap’ (use the ocular for focusing and adjust the image with min/max on the camera)
4. Go to ‘Graphics’ and select the arrow
5. Draw several lines across the pattern
6. Make sure that the ends of the arrow are placed in the same line across the image, switch off the interpolation
![Grid_left](uploads/a87bec7600fdf31fe9b0065a2f23a0f4/Grid_left.jpg){width=30%} ... ![Grid_right](uploads/662b603921d038da3c9954f49d771493/Grid_right.jpg){width=27%}
![Dimensions](uploads/b3c918277fc692ab3a6e923208c0dc5a/Dimensions.jpg){width=50%}
7. In ‘Graphics’ activate ‘M’ (measurement) and check the angle
![Grid_angle](uploads/fbea05bb3dba5378465ed6d94f36d71a/Grid_angle.jpg){width=40%}
8. Go to ‘Maintain’, and drag out the ‘Dual Camera Adapter Alignment’ tool
9. Change the ‘Rot [°]’ value. Store current position in the ‘Dual Camera Alignment’ tool
10. Take another Snap of the marker sample and measure the angle again. Repeat until you measure a value lower than 0.05°. Store current position in the ‘Dual Camera Alignment’ tool
</details>
<details>
<summary>
### `Camera xyz-Alignment`
</summary>
1. Use the 63x 1.4 (SIM) or 1.46 (TIRF) objective and the bead slide and focus on beads (z-position approx. 3.273)
2. Use the same imaging method (LASER WF, SIM, Apotome etc.) and the same filter settings as for your experiment!
![cam_z_perfect](uploads/a86788dd6a3694c136010bb0463407bf/cam_z_perfect.PNG)
3. In 'Acquisition' set up the same lasers as for your experiment: set up one laser for camera 1 (TV1) and the other laser for camera 2 (TV2) with exposure time 100ms and 1-5% of the laser power (when pressing min/max you should have >1000 white/grey values, when ticking 'Show all' in the 'Display' panel). If you use multiple tracks (in case you want to align more than two colors), always keep track 1 with camera 2!
4. Choose the optimal grid pattern
5. Go to 'Live' and focus through the beads. The optimal focus (smallest diameter of the bead) should be the same for all color channels. If not:
6. Go to ‘Maintain’ and drag out the ‘Dual Camera Adapter Alignment’ tool
7. Focus on the TV1 (camera 1) channel and change the Z value of the TV2 (camera 2) channel, until a good overlap of all color channels is achieved, especially for the beads in the center of your image. Focus up and down, and change the Z value until the beads are perfectly in focus for all color channels simultaneously. Store current position
8. Take a Z-stack (3um, center) using half of the optimal sampling (double-click on [Opt] and divide by 2)
9. SIM<sup>2</sup> process Z-Stack in case of SIM (no processing in case of TIRF or WF)
10. Go to Orthogonal View. Click on one recorded bead in the center of your image and, in 'Dual Camera Adapter Alignment tool', fine-tune the Z value if the different color channels do not overlap perfectly. Larger Z values correspond to a shift of the TV2 channel to the right in the Ortho View.
11. Store current position! Then, acquire another Z-stack and repeat until perfect overlap of all color channels is achieved.
12. In 'Live' mode change X and Y in the 'Dual Camera Adapter Alignment tool' until an XY overlap of beads in all color channels is achieved. Especially the beads in the center of the image should nicely overlap. Store current position!
13. Use the 'Profile > Display > Marker > Difference' tool to check the overlap
![cam_x_lineprofile](uploads/0b92f0c052b16846ceaaf2badf438694/cam_x_lineprofile.PNG)
</details>
<details>
<summary>
### `Channel Alignment`
</summary>
Channel Alignment has to be done before **each colocalization experiment** using multiple color channels and for **each used filter** in the camera link, for **each used objective** and for **each configuration of different cameras/tracks** individually. A bead sample has to be prepared by the user beforehand, the exact same way (same mounting etc.) as the actual sample. If your lab does not have beads, contact the facility staff and they will show you which ones to purchase (Tetraspeck 100nm beads, they are stored in the fridge's door close to the Elyra).
1. Use the 63x 1.4 (SIM) or 1.46 (TIRF) objective, use your own bead slide, and focus on beads. Use the same filters as for your experiment and set up the cameras and tracks the same way as you will use them in your experiment. Set the illumination time to 100ms and use 0.1-1% laser power (when pressing min/max you should have >1000 white/grey values, when ticking 'Show all' in the 'Display' panel). Use the optimal grid pattern.
2. Take a Z-stack (3um, center) using half of the optimal sampling (double-click on [Opt] and divide by 2)
3. SIM process your bead data
4. Go to Processing – Channel alignment
![Channel_alignment](uploads/17ba46853b6143467d8f1913ca8a3b5b/Channel_alignment.jpg){width=40%}
5. Choose your SIM processed sample-data as input image and your SIM processed beads data as input image 2, activate Fit and Affine, Click apply, save correction matrix and of course also your corrected image
</details>
## Data Processing
<details>
<summary>
### `Data Handling and Transfer`
</summary>
[Zeiss_Elyra_Data_handling_and_transfer.pdf](uploads/05626bace59dedf1f98c81eb92659ac5/Zeiss_Elyra_Data_handling_and_transfer.pdf)
</details>
<details>
<summary>
### `SIM Processing Guide`
</summary>
#### SIM processing
[UserGuide_SIM__V3.pdf](uploads/46bd2bd01d8098ac86b3ff084859b22b/UserGuide_SIM__V3.pdf)
[Zeiss_Elyra_Data_processing_manual.pdf](uploads/7eb4ca92b4533de12263da4abbd6c330/Zeiss_Elyra_Data_processing_manual.pdf)
#### SIM batch processing (multiple files at once)
[Quick_Guide_-_SIM_Batch_Processing.pdf](uploads/d1814b8e6d83d739418764b6363ce31b/Quick_Guide_-_SIM_Batch_Processing.pdf)
</details>
# Technical Specifications
[Configuration_Elyra.pdf](uploads/86be8a7457b9836c575026687c4b68c2/Configuration_Elyra.pdf)
## Objectives
```json:table
{
"fields": [
{"key": "designation", "label": "Designation", "sortable": true},
{"key": "magnification", "label": "Magnification", "sortable": true},
{"key": "na", "label": "NA", "sortable": true},
{"key": "wd", "label": "WD", "sortable": true},
{"key": "immersion", "label": "Immersion", "sortable": true},
{"key": "instrument", "label": "Instrument", "sortable": true},
{"key": "manufacturer", "label": "Manufacturer", "sortable": true},
{"key": "type", "label": "Type", "sortable": true},
{"key": "prod_num", "label": "Product Number", "sortable": true}
],
"items": [
{"designation": "EC Plan-Neofluar", "magnification": "10", "na": "0,3", "wd": "5,2", "immersion": "air", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "M27", "prod_num": "420340-9901-000"},
{"designation": "Plan-Apochromat", "magnification": "40", "na": "1,4", "wd": "0,13", "immersion": "oil", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "DIC M27", "prod_num": "420762-9900-000"},
{"designation": "C-Apochromat", "magnification": "63", "na": "1,2", "wd": "0,28", "immersion": "water", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "Corr M27", "prod_num": "421787-9970-799"},
{"designation": "Plan-Apochromat", "magnification": "63", "na": "1,4", "wd": "0,19", "immersion": "oil", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "DIC M27", "prod_num": "420782-9900-799"},
{"designation": "alpha Plan-Apochromat", "magnification": "63", "na": "1,46", "wd": "0,1", "immersion": "oil", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "Corr M27", "prod_num": "420780-9971-000"},
{"designation": "alpha Plan-Apochromat", "magnification": "100", "na": "1,46", "wd": "0,11", "immersion": "oil", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "DIC VIS M27 Airy", "prod_num": "420792-9800-720"}
]
}
<img align="right" width="100" height="73" src="uploads/9cec1f9af6048a9fafc32567a125911b/MPL-Logo-RGB-Blau-2019-05-24.jpg">
[[_TOC_]]
# Profile
| ![Elyra](uploads/f53ec5c341f61133fa78aedb7630c39e/Elyra.png){width=5000} | Inverse super-resolution microscope for lattice SIM, 3D SMLM (PALM/STORM), TIRF and laser widefield imaging. Apotome mode for optical sectioning of thick specimen. Dual-camera port with 2 sCMOS cameras for fast acquisitions. Lasers to image dyes from blue to far red. Multi-positioning (Piezo-driven XY stage with Z-Piezo insert) and hardware autofocus. Environmental box for live imaging with temperature and CO2 control. Widefield illumination: Excelitas Xylis (white LED), halogen lamp. Processing PC for offline image processing (room 1.723) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
# Manuals & Guides
## Startup & Shutdown
<details>
<summary>
### `Startup`
</summary>
1. The Elyra microscope has only one switch that needs to be turned on (Main Switch), which is a turn knob. Do not turn any other switches on/off. Next, you will need to switch on the computer under the table.
![PXL_20240320_095753421](uploads/6243a65ae74f7d9c37bec3fa7fe80a35/PXL_20240320_095753421.jpg)
2. If at any point during your experiment a loud beeping sound goes off, first check the CO<sub>2</sub> sensor (and leave the room if the sound is coming from the sensor). If not, the sound is coming from the cooling system behind the Elyra and you need to [refill the cooling liquid](Maintenance/Elyra 7#Cooling-System).
3. Choose the correct objective you want to use for imaging (via the touch panel, menu point 'Microscope') and press 'Load position' to bring the objective down to its lowest position.
![PXL_20240320_100556693](uploads/944433d431b4d0027717cf627607037c/PXL_20240320_100556693.jpg){width=49%} ![PXL_20240320_100603288](uploads/7243fdb1c0c2344cdd05237a1b72ec56/PXL_20240320_100603288.jpg){width=49%}
4. Start the 'Zen black' Software and press 'Start up System'
5. The temperature inside the incubator will be by default at 27°C to minimize drift. Use the oil for 30°C. Immersion oil bottles for 30°C and 37°C are stored in the incubator at the bottom right.
![PXL_20240320_091559891](uploads/17793d9cf5156c1cff3710c2f8ec3da2/PXL_20240320_091559891.jpg){width=50%}
6. Put on your sample and focus on it using the 'Locate' tab in the Zen Software (press 'Fluorescence' or 'Transmitted').
7. With the front window open and the light in the incubator on (switch to the left, on the microscope table) you can safely navigate larger distances when using the joystick (which has an inverse logic).
For the front flap you will need to open (to move flap up) and tighten (to keep the flap in place) the little red screw. Please handle this window with care!
8. All doors on the incubator need to be properly closed before you can start imaging in the Acquisition tab as otherwise the interlock is not released and the lasers will not be active. You can check for this: if one of the LEDs on the main controller is blinking, one of the laser interlocks has not been recognized. Unfortunately, there is no feedback which interlock is affected, so you need to check them all and potentially, open and close all doors again or even lift the condenser and put it back down.
![PXL_20240320_091545142](uploads/771b95ca313e0a923254dbf26539bca8/PXL_20240320_091545142.jpg){width=50%} ![PXL_20240320_092901039](uploads/abfc3c39fba668f2f6dfff165d7c3a91/PXL_20240320_092901039.jpg){width=48%}
</details>
<details>
<summary>
### `Shutdown`
</summary>
- Close Zen Black
- Press the Load position on the touch panel
- Clean immersion objectives with lens cleaning paper
![Cleaning](uploads/93470e14dd699b18acd95bef9b1d8c97/Cleaning.jpg){width=40%}
Leaving system for the next user
- Logout from the Windows session
Full shutdown (last user of the day)
- Shut down the PC
- Switch off the microscope
Always make sure that the system is back at 27°C and the CO<sub>2</sub> is off when you have done any live cell imaging.
</details>
## Acquisition Setup
<details>
<summary>
### `Setting up Lattice SIM`
</summary>
[Zeiss_Elyra_Setting_up_Lattice_SIM.pdf](uploads/8a812e218450d11bab8caa13642c6510/Zeiss_Elyra_Setting_up_Lattice_SIM.pdf)
</details>
<details>
<summary>
### `Setting up TIRF`
</summary>
[Zeiss_Elyra_Setting_up_TIRF_imaging.pdf](uploads/d5abb8b971c53cbf7a21b318be59d1c3/Zeiss_Elyra_Setting_up_TIRF_imaging.pdf)
</details>
<details>
<summary>
### `Environmental Control`
</summary>
[Zeiss_Elyra_Environmental_controls_01.pdf](uploads/9e4a1088f421fd1742a2dd0076f3a89a/Zeiss_Elyra_Environmental_controls_01.pdf)
</details>
## Dual-Camera Alignment & Colocalization Analysis
Camera Alignment has to be checked/done before **each experiment** using both cameras and for **each used filter** in the camera link and for **each used objective** individually. It has to be done again if the camera was pushed by accident, or if the cover of the camera link has been opened.
<details>
<summary>
### `Camera Rotation`
</summary>
Use the yellow marker pen slide with a coverslip (on the desk next to the lens cleaning paper) and the slide with beads (multispec low, they are stored in the fridge close to the Elyra).
![PXL_20240321_131615341](uploads/9c0a3a7f70adb315ded4ebd63222571c/PXL_20240321_131615341.jpg){width=40%} ![PXL_20240321_131830944](uploads/8fb5c5277bf0106adaef969aefefa79d/PXL_20240321_131830944.jpg){width=40%}
1. Select the objective that you will use for imaging
2. Set up camera 2 (TV2) for calibration (TV1 is not calibrated): lattice SIM, LBF 405/488/561/642, LP560, use the 42µm grid
![cam_rotation_perfect_63x](uploads/80b10ebcf333aae8a5d4541459184fe9/cam_rotation_perfect_63x.PNG)
3. Focus on the lattice pattern with 'Continuous' and take a ‘Snap’ (use the ocular for focusing and adjust the image with min/max on the camera)
4. Go to ‘Graphics’ and select the arrow
5. Draw several lines across the pattern
6. Make sure that the ends of the arrow are placed in the same line across the image, switch off the interpolation
![Grid_left](uploads/a87bec7600fdf31fe9b0065a2f23a0f4/Grid_left.jpg){width=30%} ... ![Grid_right](uploads/662b603921d038da3c9954f49d771493/Grid_right.jpg){width=27%}
![Dimensions](uploads/b3c918277fc692ab3a6e923208c0dc5a/Dimensions.jpg){width=50%}
7. In ‘Graphics’ activate ‘M’ (measurement) and check the angle
![Grid_angle](uploads/fbea05bb3dba5378465ed6d94f36d71a/Grid_angle.jpg){width=40%}
8. Go to ‘Maintain’, and drag out the ‘Dual Camera Adapter Alignment’ tool
9. Change the ‘Rot [°]’ value. Store current position in the ‘Dual Camera Alignment’ tool
10. Take another Snap of the marker sample and measure the angle again. Repeat until you measure a value lower than 0.05°. Store current position in the ‘Dual Camera Alignment’ tool
</details>
<details>
<summary>
### `Camera xyz-Alignment`
</summary>
1. Use the 63x 1.4 (SIM) or 1.46 (TIRF) objective and the bead slide and focus on beads (z-position approx. 3.273)
2. Use the same imaging method (LASER WF, SIM, Apotome etc.) and the same filter settings as for your experiment!
![cam_z_perfect](uploads/a86788dd6a3694c136010bb0463407bf/cam_z_perfect.PNG)
3. In 'Acquisition' set up the same lasers as for your experiment: set up one laser for camera 1 (TV1) and the other laser for camera 2 (TV2) with exposure time 100ms and 1-5% of the laser power (when pressing min/max you should have >1000 white/grey values, when ticking 'Show all' in the 'Display' panel). If you use multiple tracks (in case you want to align more than two colors), always keep track 1 with camera 2!
4. Choose the optimal grid pattern
5. Go to 'Live' and focus through the beads. The optimal focus (smallest diameter of the bead) should be the same for all color channels. If not:
6. Go to ‘Maintain’ and drag out the ‘Dual Camera Adapter Alignment’ tool
7. Focus on the TV1 (camera 1) channel and change the Z value of the TV2 (camera 2) channel, until a good overlap of all color channels is achieved, especially for the beads in the center of your image. Focus up and down, and change the Z value until the beads are perfectly in focus for all color channels simultaneously. Store current position
8. Take a Z-stack (3um, center) using half of the optimal sampling (double-click on [Opt] and divide by 2)
9. SIM<sup>2</sup> process Z-Stack in case of SIM (no processing in case of TIRF or WF)
10. Go to Orthogonal View. Click on one recorded bead in the center of your image and, in 'Dual Camera Adapter Alignment tool', fine-tune the Z value if the different color channels do not overlap perfectly. Larger Z values correspond to a shift of the TV2 channel to the right in the Ortho View.
11. Store current position! Then, acquire another Z-stack and repeat until perfect overlap of all color channels is achieved.
12. In 'Live' mode change X and Y in the 'Dual Camera Adapter Alignment tool' until an XY overlap of beads in all color channels is achieved. Especially the beads in the center of the image should nicely overlap. Store current position!
13. Use the 'Profile > Display > Marker > Difference' tool to check the overlap
![cam_x_lineprofile](uploads/0b92f0c052b16846ceaaf2badf438694/cam_x_lineprofile.PNG)
</details>
<details>
<summary>
### `Channel Alignment`
</summary>
Channel Alignment has to be done before **each colocalization experiment** using multiple color channels and for **each used filter** in the camera link, for **each used objective** and for **each configuration of different cameras/tracks** individually. A bead sample has to be prepared by the user beforehand, the exact same way (same mounting etc.) as the actual sample. If your lab does not have beads, contact the facility staff and they will show you which ones to purchase (Tetraspeck 100nm beads, they are stored in the fridge's door close to the Elyra).
1. Use the 63x 1.4 (SIM) or 1.46 (TIRF) objective, use your own bead slide, and focus on beads. Use the same filters as for your experiment and set up the cameras and tracks the same way as you will use them in your experiment. Set the illumination time to 100ms and use 0.1-1% laser power (when pressing min/max you should have >1000 white/grey values, when ticking 'Show all' in the 'Display' panel). Use the optimal grid pattern.
2. Take a Z-stack (3um, center) using half of the optimal sampling (double-click on [Opt] and divide by 2)
3. SIM process your bead data
4. Go to Processing – Channel alignment
![Channel_alignment](uploads/17ba46853b6143467d8f1913ca8a3b5b/Channel_alignment.jpg){width=40%}
5. Choose your SIM processed sample-data as input image and your SIM processed beads data as input image 2, activate Fit and Affine, Click apply, save correction matrix and of course also your corrected image
</details>
## Data Processing
<details>
<summary>
### `Data Handling and Transfer`
</summary>
[Zeiss_Elyra_Data_handling_and_transfer.pdf](uploads/05626bace59dedf1f98c81eb92659ac5/Zeiss_Elyra_Data_handling_and_transfer.pdf)
</details>
<details>
<summary>
### `SIM Processing Guide`
</summary>
#### SIM processing
[UserGuide_SIM__V3.pdf](uploads/46bd2bd01d8098ac86b3ff084859b22b/UserGuide_SIM__V3.pdf)
[Zeiss_Elyra_Data_processing_manual.pdf](uploads/7eb4ca92b4533de12263da4abbd6c330/Zeiss_Elyra_Data_processing_manual.pdf)
#### SIM batch processing (multiple files at once)
[Quick_Guide_-_SIM_Batch_Processing.pdf](uploads/d1814b8e6d83d739418764b6363ce31b/Quick_Guide_-_SIM_Batch_Processing.pdf)
</details>
# Technical Specifications
[Configuration_Elyra.pdf](uploads/86be8a7457b9836c575026687c4b68c2/Configuration_Elyra.pdf)
## Objectives
```json:table
{
"fields": [
{"key": "designation", "label": "Designation", "sortable": true},
{"key": "magnification", "label": "Magnification", "sortable": true},
{"key": "na", "label": "NA", "sortable": true},
{"key": "wd", "label": "WD", "sortable": true},
{"key": "immersion", "label": "Immersion", "sortable": true},
{"key": "instrument", "label": "Instrument", "sortable": true},
{"key": "manufacturer", "label": "Manufacturer", "sortable": true},
{"key": "type", "label": "Type", "sortable": true},
{"key": "prod_num", "label": "Product Number", "sortable": true}
],
"items": [
{"designation": "EC Plan-Neofluar", "magnification": "10", "na": "0,3", "wd": "5,2", "immersion": "air", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "M27", "prod_num": "420340-9901-000"},
{"designation": "Plan-Apochromat", "magnification": "40", "na": "1,4", "wd": "0,13", "immersion": "oil", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "DIC M27", "prod_num": "420762-9900-000"},
{"designation": "C-Apochromat", "magnification": "63", "na": "1,2", "wd": "0,28", "immersion": "water", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "Corr M27", "prod_num": "421787-9970-799"},
{"designation": "Plan-Apochromat", "magnification": "63", "na": "1,4", "wd": "0,19", "immersion": "oil", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "DIC M27", "prod_num": "420782-9900-799"},
{"designation": "alpha Plan-Apochromat", "magnification": "63", "na": "1,46", "wd": "0,1", "immersion": "oil", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "Corr M27", "prod_num": "420780-9971-000"},
{"designation": "alpha Plan-Apochromat", "magnification": "100", "na": "1,46", "wd": "0,11", "immersion": "oil", "instrument": "Elyra-7", "manufacturer": "Zeiss", "type": "DIC VIS M27 Airy", "prod_num": "420792-9800-720"}
]
}
```
\ No newline at end of file