@@ -156,7 +156,24 @@ Camera Alignment has to be checked/done before **each experiment** using both ca
</summary>
hi
-
Use or set up a new Track in which both cameras are used. Click on DuoLink and click on "Opens alignment tool".

{width=481 height=738}
-
In the new Dual Camera Alignment window, tick "Use alignment target" and start Live acquisition.

{width=1065 height=1067}
-
In the Dual Camera Alignment window, click on XY button next to Adjustment. This will automatically align the two cameras in X and Y directions.

{width=1524 height=989}-
-
The alignment might not be perfect, and you might need to adjust the X and Y positions manually with the sliders.
-
If one of the channels seems more defocused then the other, adjust the Z direction with the slider.
-
Only change the Rotation if the L shaped object in the middle can not be aligned with changing X and Y directions.
-
At the end, click "Store Calibration" button
</details>
...
...
...
...