...@@ -156,7 +156,24 @@ Camera Alignment has to be checked/done before **each experiment** using both ca ...@@ -156,7 +156,24 @@ Camera Alignment has to be checked/done before **each experiment** using both ca
</summary> </summary>
hi - Use or set up a new Track in which both cameras are used. Click on DuoLink and click on "Opens alignment tool".
![image](uploads/cd54dd3b8dce81909fa33a3c3b019e30/image.png){width=481 height=738}
- In the new Dual Camera Alignment window, tick "Use alignment target" and start Live acquisition.
![image](uploads/39ac4c072fb82af56e4f0ea1c98e2c3d/image.png){width=1065 height=1067}
- In the Dual Camera Alignment window, click on XY button next to Adjustment. This will automatically align the two cameras in X and Y directions.
![image](uploads/b55bd605e76329b64633567626c6e717/image.png){width=1524 height=989}-
- The alignment might not be perfect, and you might need to adjust the X and Y positions manually with the sliders.
- If one of the channels seems more defocused then the other, adjust the Z direction with the slider.
- Only change the Rotation if the L shaped object in the middle can not be aligned with changing X and Y directions.
- At the end, click "Store Calibration" button