Update Zeiss Elyra 7 authored by Clara Bodner's avatar Clara Bodner
......@@ -29,12 +29,12 @@ Use a yellow marker pen slide with a coverslip and a slide with beads (100nm Tet
5. Make sure that the ends of the arrow are placed in the same line across the image, switch off the interpolation
![Grid_left](uploads/a87bec7600fdf31fe9b0065a2f23a0f4/Grid_left.jpg){width=30%} ![Grid_right](uploads/662b603921d038da3c9954f49d771493/Grid_right.jpg){width=27%}
![Dimensions](uploads/b3c918277fc692ab3a6e923208c0dc5a/Dimensions.jpg){width=30%}
![Grid_left](uploads/a87bec7600fdf31fe9b0065a2f23a0f4/Grid_left.jpg){width=30%} ... ![Grid_right](uploads/662b603921d038da3c9954f49d771493/Grid_right.jpg){width=27%}
![Dimensions](uploads/b3c918277fc692ab3a6e923208c0dc5a/Dimensions.jpg){width=35%}
6. In ‘Graphics’ activate ‘M’ (measurement) and check the angle
![Graphics_measure](uploads/8de9fc67bee34262465186b8022c2c1f/Graphics_measure.jpg){width=30%}
![Graphics_measure](uploads/8de9fc67bee34262465186b8022c2c1f/Graphics_measure.jpg){width=35%}
![Grid_angle](uploads/fbea05bb3dba5378465ed6d94f36d71a/Grid_angle.jpg){width=30%}
7. Go to ‘Maintain’, and drag out the ‘Dual Camera Adapter Alignment’ tool
......
......