## Environmental Control
|
|
|
|
|
|
- Stagetop incubation only
|
|
|
- Controller: Zeiss TempModule S, CO<sub>2</sub> Module S
|
|
|
|
|
|
## Acquisition Setup
|
|
|
|
|
|
|
|
|
<details>
|
|
|
|
|
|
<summary>
|
|
|
|
|
|
### `Setting up Incubation`
|
|
|
### `Environmental Control`
|
|
|
|
|
|
</summary>
|
|
|
|
|
|
- Stagetop incubation only
|
|
|
- Controller: Zeiss TempModule S, CO<sub>2</sub> Module S
|
|
|
|
|
|
If you want plan to use stage-top incubation on the LSM-980, you need to first open (only one turn needed) the valve on the CO<sub>2</sub> bottle itself, then open the respective valve to the microscope.
|
|
|
|
|
|
Next, assemble the heating unit with the appropriate insert for your sample carrier. There is one insert for regular slides/Ibidi Dishes **(3)** and one for 35mm dishes **(4)**
|
... | ... | @@ -88,8 +91,6 @@ In the software, move to the "Incubation" Tab (in the right hand panel in ZEN) a |
|
|
|
|
|
</details>
|
|
|
|
|
|
## Acquisition Setup
|
|
|
|
|
|
The LSM-980 can be used in various modes, the "standard" confocal mode on the one hand and a variety of imaging modalities using the Airyscan detector. Depending on the modalities selected, setting up your acquisition will vary slightly and different aspects need to be considered.
|
|
|
|
|
|
In both cases the "Smart Setup" option is a good point to start from.
|
... | ... | |