### Environmental control (stage top)
|
|
|
|
|
|
- Stagetop incubation only
|
|
|
- Controller: Zeiss TempModule S, CO2 Module S
|
|
|
- Controller: Zeiss TempModule S, CO<sub>2</sub> Module S
|
|
|
|
|
|
If you want plan to use stage-top incubation on the LSM-980, you need to first open (only one turn needed) the valve on the CO2 bottle itself, then open the respective valve to the microscope.
|
|
|
If you want plan to use stage-top incubation on the LSM-980, you need to first open (only one turn needed) the valve on the CO<sub>2</sub> bottle itself, then open the respective valve to the microscope.
|
|
|
|
|
|
Next, assemble the heating unit with the appropriate insert for your sample carrier. There is one insert for regular slides/Ibidi Dishes **(3)** and one for 35mm dishes **(4)**
|
|
|
|
... | ... | @@ -61,7 +61,7 @@ Finally add the heated lid - it should fit exactly over the heated insert. The f |
|
|
|
|
|
**Be careful when lowering the condenser, as it could be in a too low position, which would lead to the condenser crashing into the heating lid!**
|
|
|
|
|
|
In the software, move to the "Incubation" Tab (in the right hand panel in ZEN) and tick the heated lid and one of the inserts (the one you use) and the components will be heated to the selected temperature. Finally tick "CO<sub>2</sub> Small V", assuming you need to supply Co2.
|
|
|
In the software, move to the "Incubation" Tab (in the right hand panel in ZEN) and tick the heated lid and one of the inserts (the one you use) and the components will be heated to the selected temperature. Finally tick "CO<sub>2</sub> Small V", assuming you need to supply CO<sub>2</sub>.
|
|
|
|
|
|
**Inc PM**: lid
|
|
|
**H Insert P**: insert for slides/ibidi dishes
|
... | ... | |